The case
As part of a partnership with CARMAT and the Laboratory of Fluid Mechanics and Acoustics (LMFA), Dynas+ developed a numerical fluid-structure interaction (FSI) model to reproduce the behavior of the hyperelastic membrane of a total artificial heart. The goal was to correlate the LS-DYNA simulation results with experimental measurements taken on an instrumented test bench, in order to validate the model's ability to simultaneously reproduce membrane deformation and the pulsatile flows generated within the device.
The solution deployed
Dynas+ developed a high-fidelity FSI model in LS-DYNA, combining the ICFD solver with a hyperelastic membrane model calibrated from mechanical tests provided by CARMAT. The methodology implemented includes:
- calibration of the membrane's hyperelastic behavior through test/simulation correlation;
- generation of the fluid and structure models in DEP MeshWorks;
- strong coupling between the incompressible flows and the deformable membrane;
- reproduction of the actual piston motion of the test bench;
- quantitative comparison of the numerical results with PIV and 4D-PTV measurements.

The results achieved
The developed model reproduces the overall behavior of the experimental bench with a good level of fidelity. The main results show:
- excellent correlation between the simulated membrane deformations and the experimental measurements;
- faithful reproduction of the velocity fields measured by PIV, with limited deviations related to the modeling assumptions;
- a maximum membrane deformation reaching approximately 95% of the experimental value;
- validation of LS-DYNA's ability to simulate strongly coupled fluid-structure interactions in the field of medical devices.
This work was the subject of a scientific paper presented at the 2025 LS-DYNA Conference (Munich): "Fluid Structure Interaction Modeling of an Artificial Heart Membrane Test Bench".